Catalog Number 196000400
Medical Device Problem Codes Use of Device Problem (1670); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 01/01/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Article entitled ¿3d polyethylene wear analysis after primary cruciate-retaining tka failure using optical scanning and cad models references¿ written by matej valic, vesna leva¿ic, ingrid milo¿ev, mateja blas, eva podov¿ovnik, jaka koren, and rihard treb¿e was reviewed.Our purpose was to identify (a) whether volumetric wear (adjusted for time in-situ) significantly differ among groups of common causes of tka failure, (b) whether there is a characteristic wear distribution pattern for a common cause of tka failure, and (c) whether nominal insert size (small vs large) and components size ratio (femur-to-insert) influence on volumetric wear rates.The study included patients that had been implanted with a sigma tka (cemented tibial tray and porous coated femoral component).Within the patient that met the criteria, there were noted to be 86 revision cases due to failure (no mention of why).59/86 tibial inserts were retrieved.2 of the 59 inserts exhibited inexplicable abnormal type of damage and were excluded from the final analysis.Final cohort included 57 inserts.There was no mention of cement mfg or patella implants.Results: 86 revisions.Only 5 patients had a single detected causes, while the remaining had 2-3 detected causes.35 cases of infection (13 of the 35 cases were noted to have loosening as well), 18 cases of osteolysis and loosening (unknown device), 6 cases of osteolysis, 28 cases of joint malignment or component malpositioning, 12 cases of instability, 10 cases with ligament insufficiency, and 3 cases for other isolated causes.Volumetric and linear wear noted on the 57 liners.2 liners excluded from analysis are pictured in figure 5 ¿ show signs of wear.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC SIG CR NPOR FEM LT SZ 4N
Common Device Name
SIGMA KNEE PRIMARY : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key19172266
Report Number1818910-2024-08927
Device Sequence Number12534020
Product Code JWH
UDI-Device Identifier10603295074762
UDI-Public10603295074762
Combination Product (Y/N)N
Initial Reporter CountrySI
PMA/510(K) Number
K943462
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Literature,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number196000400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/14/2024
Supplement Date Received by Manufacturer04/30/2024
Initial Report FDA Received Date04/24/2024
Supplement Report FDA Received Date04/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-