• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSSIO LTD. OSSIOFIBER COMPRESSION STAPLE SYSTEM 20X20MM; STAPLE, ABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSSIO LTD. OSSIOFIBER COMPRESSION STAPLE SYSTEM 20X20MM; STAPLE, ABSORBABLE Back to Search Results
Model Number OF2062020S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Local Reaction (2035)
Date of Event 03/26/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Company was notified of a scheduled removal surgery following localized pain in the operated left foot a year after implantation.The removal surgery was completed successfully.An investigation was performed including internal analysis of records, patient imaging, lab work and surgeon feedback.There is no indication of a design, manufacturing, sterilization, or process issue affecting implant safety or effectiveness.There were no issues during the primary procedure.Patient post op recovery was uneventful, and bone healing and fusion of the operated sites was achieved.Imaging due to the reported pain revealed localized bone edema.Infection was ruled out by the lab work, negative culture, and timing of symptoms.The investigation concluded that the root cause of the pain could not be determined but is likely the outcome of the patient experiencing a local reaction to one of the implant's materials or related to an unreported underlying medical condition.Instructions for use list an allergic reaction as a potential adverse effect to implantation of foreign material.Company continues to monitor these events as part of post market activities.Additional reports relating to this event: # 3014554088-2024-00003/4/5/6 & 3014323288-2024-00003/4/5.
 
Event or Problem Description
Pain and localized bone edema in the left foot following a bunion deformity correction.Four implants were removed one year post-op.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSIOFIBER COMPRESSION STAPLE SYSTEM 20X20MM
Common Device Name
STAPLE, ABSORBABLE
Manufacturer (Section D)
OSSIO LTD.
8 hatochen st.
caesarea, 30798 61
IS  3079861
MDR Report Key19173193
Report Number3014323288-2024-00006
Device Sequence Number15438478
Product Code MNU
UDI-Device Identifier07290017630700
UDI-Public(01)07290017630700(17)240416(11)221218(10)OF02229
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/24/2024
Report Date (Section F) 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/16/2024
Device Model NumberOF2062020S
Device Catalogue NumberOF2062020S
Device Lot NumberOF02229
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date03/27/2024
Device Age3 MO
Event Location Ambulatory Surgical Facility
Date Report to Manufacturer03/27/2024
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date04/24/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OSSIOFIBER COMPRESSION SCREW 3.5X22MM; OSSIOFIBER COMPRESSION SCREW 3.5X24MM; OSSIOFIBER COMPRESSION SCREW 4.0X38MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age19 YR
Patient SexFemale
Date Report Sent to FDA04/24/2024
-
-