| Model Number |
FIBERTAK DISPOSABLES STRAIGHT KIT |
| Medical Device Problem Code |
Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the drill bit seized in the guide during use in an mpfl case.They used another ar-3638ds to complete the case.
|
| |
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint confirmed.One unpackaged ar-3638ds drill guide was received for investigation.Visual inspection identified that a drill bit was seized within the guide and attempts to separate the two devices were unsuccessful.A slight bend in the guide shaft was noted.The observed condition is most likely the result of prying/leveraging the bit within the guide, resulting in the two devices binding.
|
| |
|