| Model Number |
ESBF2314C103E |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
An endurant iis stent graft system was implanted during the endovascular treatment of a penetrating aortic ulcer (pau) of the infra-renal aorta.The proximal aortic neck diameter measured 19mm proximally <(>&<)> 19mm distally with a neck length of 40mm.Infrarenal angulation was 28 degrees.Mild neck calcification was present.Mild bilateral iliac calcification was noted.The celiacwas enlarged at 16.8mm and the distal lcfa also enlarged at 15.7mm due to the fact treating a pau minimal oversizing in cia's sealing with 13mm limbs was recommended.It was reported at the one-month follow-up on 23-apr-2024, ct imaging identified mild infolding just above the flow divider of esbf2314c103e.No endoleaks were detected, and the pau showed no signs of expansion/ filling.Per the physician the cause the infolding is undetermined.No additional clinical sequelae were reported, and the patient will be monitored.
|
| |
|
Additional Manufacturer Narrative
|
|
Film evaluation summary: the reported stent graft infolding was confirmed on the films provided; however, the cause of the event could not be determined.There was no excessive oversizing that could have been attributed to the infolding.Analysis of the returned films did not reveal any out of specification stent graft integrity issues.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis conclusion: a slight infold was still observed on post-implant cts taken approximately 8.5 months post-index procedure.There is no endoleak or graft thrombus noted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|