• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC PENTARAY NAV HIGH-DENSITY MAPPING ECO CATHETER; CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC PENTARAY NAV HIGH-DENSITY MAPPING ECO CATHETER; CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY Back to Search Results
Catalog Number D128210
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/29/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.E1.Initial reporter phone: (b)(6).If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.#(b)(4).
 
Event or Problem Description
It was reported that a patient underwent an atrial flutter ablation procedure with a pentaray nav high-density mapping eco catheter and the deflection became stuck.It was reported that during the operation, catheter was unable to deflect or relax completely.A second device was used to complete the operation.There was no adverse event reported on patient.On 7-apr-2024, additional information was received indicating the catheter was stuck in a deflected position.The knob was unable to be turned or pushed down, but there was no difficulty removing the catheter.The catheter was being used with a synaptic 8.5f long sheath.This event was originally considered non-reportable, however, bwi became aware of additional information on 7-apr-2024 and reassessed the event as mdr reportable.
 
Additional Manufacturer Narrative
On 15-may-2024, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Additional Manufacturer Narrative
It was reported that a patient underwent an atrial flutter ablation procedure with a pentaray nav high-density mapping eco catheter and the deflection became stuck.It was reported that during the operation, catheter was unable to deflect or relax completely.A second device was used to complete the operation.There was no adverse event reported on patient.On 7-apr-2024, additional information was received indicating the catheter was stuck in a deflected position.The knob was unable to be turned or pushed down, but there was no difficulty removing the catheter.The catheter was being used with a synaptic 8.5f long sheath.Device evaluation details: the device was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.A visual inspection and deflection test of the returned device were performed following bwi procedures.Visual analysis revealed no damage or anomalies on the device.A deflection test was performed, and the curve was deflecting within specifications.The mechanism worked correctly, no stuck was identified during the analysis.No deflection issues were observed.A manufacturing record evaluation was performed for the finished device batch number, and no internal action were identified.The stuck issue reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the usage of the device that compromised its performance.The instructions for use contain (ifu) the following warning and precaution: always pull the thumb knob back to straighten the catheter tip before insertion or withdrawal of the catheter.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTARAY NAV HIGH-DENSITY MAPPING ECO CATHETER
Common Device Name
CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key19175670
Report Number2029046-2024-01353
Device Sequence Number3728059
Product Code MTD
UDI-Device Identifier10846835012248
UDI-Public10846835012248
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K123837
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/14/2026
Device Catalogue NumberD128210
Device Lot Number31069503L
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/07/2024
Supplement Date Received by Manufacturer05/15/2024
05/23/2024
Initial Report FDA Received Date04/24/2024
Supplement Report FDA Received Date05/16/2024
06/14/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SYNAPTIC 8.5F LONG SHEATH
Patient SexUnknown
-
-