| Model Number |
8886621741-2 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during a laparotomy for a veterinary procedure, when the package was opened, the needle and thread detached four times in a row during normal suturing.A new suture was used to resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: 8886621741-2, maxon 3-0 30 grn cv-24 8886621741-2 (lot#: d3a1729y); 8886621741-2, maxon 3-0 30 grn cv-24 8886621741-2 (lot#: d3a1729y); 8886621741-2, maxon 3-0 30 grn cv-24 8886621741-2 (lot#: d3a1729y); 8886621741-2, maxon 3-0 30 grn cv-24 8886621741-2 (lot#: d3a1729y) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the needle was returned disengaged from the suture.Upon microscopic inspection of the needle, normal crimp and swage marks were noted.Functional testing found that a needle attachment test was performed on sealed representative samples and the results met the specifications for this product.It was reported that the needle detached.The reported issue was confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition: needle bent.Visual examination noted that the needle was bent near the swaged end of the needle.Instrument marks were observed at the bend site.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|