| Catalog Number |
33680034 |
| Medical Device Problem Codes |
Osseointegration Problem (3003); Migration (4003)
|
| Health Effect - Clinical Codes |
Inadequate Osseointegration (2646); Implant Pain (4561); Swelling/ Edema (4577)
|
| Date of Event |
12/18/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.H3 other text : device remains implanted in patient.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented in the clinic with complaints of pain and swelling on the left ankle.Upon physical examination showed medial lateral left ankle tenderness to palpitation with mid left ankle swelling.The anterior wound is well-healed with no erythema or sign of infection.X-rays reviewed indicated that there may some talar subsidence and loosening.No obvious fractures were evident.It was recommend that the patient continue home exercise program and use prescribed voltaren gel.A spect ct was ordered to evaluate any loosing in subsidence.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event that could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented in the clinic with complaints of pain and swelling on the left ankle.Upon physical examination showed medial lateral left ankle tenderness to palpitation with mid left ankle swelling.The anterior wound is well-healed with no erythema or sign of infection.X-rays reviewed indicated that there may some talar subsidence and loosening.No obvious fractures were evident.It was recommend that the patient continue home exercise program and use prescribed voltaren gel.A spect ct was ordered to evaluate any loosing in subsidence.
|
| |
|