| Catalog Number |
367820 |
| Medical Device Problem Code |
Device Ingredient or Reagent Problem (2910)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: bd had not received samples or photos for investigation.Therefore, 30 retention samples from bd inventory were evaluated by visual examination and no issues were observed relating to poor serum as all samples met specifications.Complaints for sample quality are under statistical control for the month of april 2024.At this time, further testing is not indicated.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode poor serum.Bd was not able to identify an exact root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.Bd quality will continue to monitor sample quality complaints.H3 other text : see h10.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® serum blood collection tubes that there were red blood cells in the serum after centrifugation resulting in poor serum; inadequate separation of the sample.There was no report of impact on the patient or user.
|
| |
|