Section h10: (h3) pending evaluation.(d10) concomitant device(s): 531-78-20 - shouldr gps hex pins kit 6265431.A10012 - gps implant kit v2 10000219262.300-01-09 - equinoxe, humeral stem primary, press fit 9mm 6119646.300-20-02 - equinox square torque define screw drive kit 6104214.300-50-15 - 1.5mm short rep plate 5293116.310-00-47 - xs humeral head, 47mm xs offset humeral head 5289479.
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: g1, g3, h6, h11 mdr section codes updated/corrected: a, b, c, d, e, g the revision reported was likely the result of prosthesis fracture of the center cage leading to fracture, wear, and disassembly of the glenoid component.The cause of the center cage fracture cannot be confirmed but may be due to improper initial seating of the glenoid and/or due to off-axis drilling of the peripheral pegs resulting in the rocking horse phenomenon.However, the cause and series of events cannot be confirmed as the devices were not returned for evaluation, and initial post-operative radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|