• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE CAGE GLENOID MEDIUM, ALPHA; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE CAGE GLENOID MEDIUM, ALPHA; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 314-13-03
Medical Device Problem Codes Fracture (1260); Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 03/28/2024
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, the 71 year old female patient had an initial right tsa on (b)(6) 2019.The patient was revised on (b)(6) 2024 due to a failed glenoid.There was no reported breakage of a device or surgical delay/prolongation.The patient was last known to be in stable condition following the event.
 
Additional Manufacturer Narrative
Section h10: (h3) pending evaluation.(d10) concomitant device(s): 531-78-20 - shouldr gps hex pins kit 6265431.A10012 - gps implant kit v2 10000219262.300-01-09 - equinoxe, humeral stem primary, press fit 9mm 6119646.300-20-02 - equinox square torque define screw drive kit 6104214.300-50-15 - 1.5mm short rep plate 5293116.310-00-47 - xs humeral head, 47mm xs offset humeral head 5289479.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: g1, g3, h6, h11 mdr section codes updated/corrected: a, b, c, d, e, g the revision reported was likely the result of prosthesis fracture of the center cage leading to fracture, wear, and disassembly of the glenoid component.The cause of the center cage fracture cannot be confirmed but may be due to improper initial seating of the glenoid and/or due to off-axis drilling of the peripheral pegs resulting in the rocking horse phenomenon.However, the cause and series of events cannot be confirmed as the devices were not returned for evaluation, and initial post-operative radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE CAGE GLENOID MEDIUM, ALPHA
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key19177881
Report Number1038671-2024-00972
Device Sequence Number17889757
Product Code KWT
UDI-Device Identifier10885862172686
UDI-Public10885862172686
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K113309
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/01/2023
Device Catalogue Number314-13-03
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2024
Supplement Date Received by Manufacturer09/02/2025
Initial Report FDA Received Date04/24/2024
Supplement Report FDA Received Date09/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/02/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexFemale
-
-