EXACTECH, INC. EQUINOXE CAGE GLENOID LARGE, BETA; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Catalog Number |
314-13-14 |
| Medical Device Problem Codes |
Crack (1135); Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
04/01/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, the 86 year old male patient had an initial right tsa in (b)(6) 2016 or (b)(6) 2015.The patient was revised on (b)(6) 2024 due to wear and tear.The cage glenoid was worn out and was loose in the joint with only the cage left, which was safely removed.The poly implant was cracked and had a significant amount of cement nearly covering the back of the implant.No fragments, parts or pieces fell into the patient.Implant and pegs all removed.There was no reported surgical delay/prolongation.The patient was last known to be in stable condition following the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.Serial number, expiration and manufactured dates unknown.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, loosening, and/or cracking.Potential contributions of manufacturing, user, and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, relevant clinical information, and product information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing data could not be reviewed for the anatomic glenoid component because serial number(s) were not provided, and the original surgery invoice could not be located from a search of exactech¿s surgery data.User-related considerations: there were no user-related issues reported that appear to have contributed to the reported event.Patient-related considerations: there are no patient conditions reported that appear to have contributed to the reported event.The reason for the revision reported in (b)(4) cannot be confirmed from the information provided but may be the result of prosthesis wear, loosening, and/or cracking.Potential contributions of manufacturing, user, and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, relevant clinical information, and product information were not provided.
|
| |
|