U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA GMK-SPHERE TIBIAL TRAY FIXED CEMENTED SIZE 4 L; KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA GMK-SPHERE TIBIAL TRAY FIXED CEMENTED SIZE 4 L; KNEE TIBIAL TRAY Back to Search Results
Catalog Number 02.07.1204L
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Medical device site infection (4845)
Date of Event 03/27/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 02-apr-2024.Lot 168504: 99 items manufactured and released on 29-mar-2017.Expiration date: 2022-03-21.No anomalies found related to the problem.To date, 95 items of the same lot have been sold with no similar reported event during the period of review.Additional components involved: gmk-sphere 02.12.0024l femoral component sphere cemented size 4+ l (k1408269) lot 163999: (b)(4) items manufactured and released on 22-sep-2016.Expiration date: 2021-09-07.No anomalies found related to the problem.To date, all the items of the same lot have been sold with no similar reported event during the period of review.Gmk-sphere 02.12.0414fl tibial insert fixed sphere flex size 4/14 mm l (k121416) lot 144605: (b)(4) items manufactured and released on 24-sep-2014.Expiration date: 2019-08-31.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.
 
Event or Problem Description
At about 6 years and 7 months after primary, the patient came in due to signs of an infection and the pathogen is unknown.The surgeon performed a washout and revised all components to hinge components.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK-SPHERE TIBIAL TRAY FIXED CEMENTED SIZE 4 L
Common Device Name
KNEE TIBIAL TRAY
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key19178382
Report Number3005180920-2024-00243
Device Sequence Number7298633
Product Code JWH
UDI-Device Identifier07630030819902
UDI-Public07630030819902
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K090988
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/21/2022
Device Catalogue Number02.07.1204L
Device Lot Number168504
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2024
Initial Report FDA Received Date04/25/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/29/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient SexMale
Patient Weight95 KG
-
-