| Model Number |
20E |
| Medical Device Problem Code |
Energy Output Problem (1431)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker requested additional information on the patient from the customer.The customer informed stryker that no further patient information is available.Patient fields in which information was not provided were intentionally left blank.Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that they were using their device on a patient and the device only delivered a shock of 79 joules at a selected energy level of 200 joules.There were no reports of any adverse effects to the patient as a result of the reported issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and was unable to duplicate the reported issue.Stryker downloaded the device's electronic records from the event for review and was able to verify the reported issue.Stryker replaced the device's therapy pcb assembly as a precaution.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that they were using their device on a patient and the device only delivered a shock of 79 joules at a selected energy level of 200 joules.There were no reports of any adverse effects to the patient as a result of the reported issue.
|
| |
|