Model Number 20E
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/27/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Stryker requested additional information on the patient from the customer.The customer informed stryker that no further patient information is available.Patient fields in which information was not provided were intentionally left blank.Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event or Problem Description
The customer contacted stryker to report that they were using their device on a patient and the device only delivered a shock of 79 joules at a selected energy level of 200 joules.There were no reports of any adverse effects to the patient as a result of the reported issue.
 
Additional Manufacturer Narrative
Stryker evaluated the customer's device and was unable to duplicate the reported issue.Stryker downloaded the device's electronic records from the event for review and was able to verify the reported issue.Stryker replaced the device's therapy pcb assembly as a precaution.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
 
Event or Problem Description
The customer contacted stryker to report that they were using their device on a patient and the device only delivered a shock of 79 joules at a selected energy level of 200 joules.There were no reports of any adverse effects to the patient as a result of the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 20E DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
brian blakeslee
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key19178646
Report Number0003015876-2024-01240
Device Sequence Number17992781
Product Code MKJ
UDI-Device Identifier00883873813829
UDI-Public00883873813829
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P160026
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number20E
Device Catalogue Number99507-000016
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2024
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/02/2024
Supplement Date Received by Manufacturer08/19/2024
Initial Report FDA Received Date04/25/2024
Supplement Report FDA Received Date09/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/21/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-