| Model Number |
NANONEEDLE DIAGNOSTIC SHEATH KIT, 125MM |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a hand joint arthroscopy the trocar did not hold and slipped out.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with the same device.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.One flexible obturator from an ar-3210-0050 nanoneedle scope diagnostic sheath kit was received for investigation.Visual evaluation noted that the flexible obturator was bent.Functional testing was unable to be performed due to the missing components.The most likely cause for the reported failure can be attributed to misuse due to the damage to the device.Refer to investigation photos.
|
| |
|