| Catalog Number |
07K65-39 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Completed information for sections a1 patient identifier and a2 age at time of event/age units: sid (b)(6) for 65 year old and sid (b)(6) for 86 year old an evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely elevated architect free t4 results on two patients that were not reported out of the lab.The following results were provided (customer provided reference range: 0.70 to 1.48 ng/ml): 65 year old: sid (b)(6) , initial result = 2.89 ng/dl, repeats = 1.58 ng/dl and 1.40 ng/ml additional laboratory data was provided: tsh = 2.08 uiu/ml 86 year old: sid (b)(6) , initial result = 3.70 ng/dl, repeats = 2.13 ng/dl, 1.48 ng/dl, and 1.32 ng/dl additional laboratory data was provided: tsh = 1.94 uiu/ml there was no impact to patient management reported.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely elevated architect free t4 results on two patients that were not reported out of the lab.The following results were provided (customer provided reference range: 0.70 to 1.48 ng/ml): 65 year old: sid (b)(6) initial result = 2.89 ng/dl, repeats = 1.58 ng/dl and 1.40 ng/ml additional laboratory data was provided: tsh = 2.08 uiu/ml.86 year old: sid (b)(6) initial result = 3.70 ng/dl, repeats = 2.13 ng/dl, 1.48 ng/dl, and 1.32 ng/dl additional laboratory data was provided: tsh = 1.94 uiu/ml.There was no impact to patient management reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Return testing was not completed as returns were not available.Data and information provided by the customer were reviewed and support the complaint issue.An increase in complaints has been observed for lot 56001ud01 and 56001ud, however, in-house performance testing was completed which indicates the product is performing as expected.Device history record review did not identify any non-conformances, potential non-conformances, or deviations associated with lot 56001ud01 and the complaint issue.Labeling was reviewed and sufficiently addresses the customer¿s issue.Based on the investigation, no systemic issue or deficiency with the architect free t4 for reagent for lot 56001ud01 was identified.
|
| |
|
Search Alerts/Recalls
|