UNOMEDICAL S.R.O. GENTLECATH OTHER; CATHETER, UROLOGICAL Back to Search Results
Model Number 501003
Medical Device Problem Code Sharp Edges (4013)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6) based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 3005778470.
 
Event or Problem Description
End user states "some of the eyelets on his catheters sharp and painful during insertion and removal.".
 
Additional Manufacturer Narrative
Review of the dhr showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.The process parameters were adjusted within the validated window.No nonconformity has been registered during the manufacturing process of the affected lot and of the mentioned issue.No similar complaint was received on mention lot#3c00806.Percentage of affected pieces against lot is (b)(4).This complaint was first presented to the complaint review board on may 16, 2024, and the investigation started.On may 20, 2024, the attendees decided that no capa is needed.Based on provided information, we cannot consider the sharpness of the eyelet.Operators will be retrained according to g708002 education and training of production personnel and the issue will be presented to operators according to (b)(4) qa data visualization.Affected quantity is within aql 0,65.Based on the investigation results the issue is considered to be isolated.Should additional information become available, a follow-up report will be submitted.Fda registration number, reporting site: 1049092, manufacturing site: 3005778470.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GENTLECATH OTHER
Common Device Name
CATHETER, UROLOGICAL
Manufacturer (Section D)
UNOMEDICAL S.R.O.
priemyselny park 3,
michalovce 07101
LO  07101
Manufacturer (Section G)
UNOMEDICAL S.R.O.
priemyselny park 3,
michalovce 07101
LO   07101
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 27409
3365424679
MDR Report Key19181038
Report Number3005778470-2024-00848
Device Sequence Number7298720
Product Code KOD
UDI-Device Identifier00768455145461
UDI-Public00768455145461
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K172924
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number501003
Device Lot Number3C00806
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2024
Supplement Date Received by Manufacturer05/22/2024
Initial Report FDA Received Date04/25/2024
Supplement Report FDA Received Date05/24/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/29/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-