• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION AED PLUS; DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION AED PLUS; DEFIBRILLATOR Back to Search Results
Model Number AED PLUS
Medical Device Problem Code Failure of Device to Self-Test (2937)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Complainant alleged that during functional testing, the device prompted a "unit failed" message.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Additional Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including bench handling, impedance, and defibrillation cycling without duplicating the report.The hv capacitor was replaced as a precaution.The device was recertified and returned to manufacturing and a replacement was sent to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AED PLUS
Common Device Name
DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key19181221
Report Number1220908-2024-01436
Device Sequence Number17949998
Product Code MKJ
UDI-Device Identifier04560233596016
UDI-Public04560233596016
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K011541/P160
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberAED PLUS
Device Catalogue NumberAED PLUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/25/2024
Supplement Report FDA Received Date05/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-