| Model Number |
PT-000435 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
According to the complainant the device will not be returned for investigation.We are unable to confirm the reported needle mechanism failure or to determine its root cause.Lot release records were reviewed and the product lot met all acceptance criteria.Specifically, a pod is paired to a pdm and put through simulated use testing including communicating with the pdm, deployment, delivering fluid, occlusion detection, and freedom from hazard alarms.
|
| |
|
Event or Problem Description
|
|
It was reported by the patient that the pod's cannula did not deploy indicating a needle mechanism failure.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received not deployed.Rotational sensor errors were seen in the download data causing first prime to end earlier than expected and the firing flat not being properly lined up by the end of second prime.Rotational sensor errors were replicated in the rerun data.Contamination was observed on the commutator cap, which can be confirmed as the cause of the rotational sensor errors and device not deploying.
|
| |
|