| Model Number |
DBL LOADED 4.75MM PEEK SWVLK |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported during a rotator cuff repair procedure, the surgeon inserted and seated the ar-2324pslc-2 double loaded peek swivelock, (lot: 12900335) as the surgeon was tightening down on the white suture, he notice the suture was cut from the eyelet.The broken suture was removed from the patient.The blue suture was use to secure the ar-2324pslc-2.The surgeon used a second ar-2324pslc-2 (lot: 13262585) and the same issue occurred.The case was completed without patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint device was not received for evaluation.Based off the information provided, the most likely cause for the reported failure can be attributed to user error of the device due to excessive force being used when tensioning the sutures.
|
| |
|