| Catalog Number |
MS26372 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation has just started; results will be provided in a follow-up report.
|
| |
|
Event or Problem Description
|
|
It was reported that on standalone m540, the spo2 waveform and numeric values dropped on (b)(6).At 10:30 am while using masimo rainbow pods on a neonatal patient.It was reported that the m540 would display that it was "searching" for an sp02 value, but that the waveform would flatline and no numerical values could be shown.No patient injury or death was reported.
|
| |
|
Event or Problem Description
|
|
It was reported that on standalone m540, the spo2 waveform and numeric values dropped on 03/28 at 10:30 am while using masimo rainbow pods on a neonatal patient.It was reported that the m540 would display that it was "searching" for an sp02 value, but that the waveform would flatline and no numerical values could be shown.No patient injury or death was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer reported that the spo2 waveform and numeric values drop intermittently at the customer site in multiple rooms.The provided m540 logs were reviewed but showed the device was in standby mode during the date/time specified (february 28th at 10:30am).Images and/ or videos of the error were requested multiple times, but not provided.Therefore, the root cause nor the issue can be confirmed.
|
| |
|