| Catalog Number |
24-1301-22 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H3 other: device has not been returned to date.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the connector on the device completely snapped off.There was patient involvement and no adverse event/human harm.
|
| |
|
Additional Manufacturer Narrative
|
|
D9.Date returned to mfg: 09-may-2024.H6.Investigation codes: updated.Investigation summary: one used decontaminated sample of 007/374/418, which is part of the 24-1301-22 nrfit connector, and one sample of lockit plus - which isn´t part of claimed 24-1301-22 - were received without their original packaging for evaluation.Received samples were visually inspected and there was no problem found.According to the event description the claimed issue was observed at the connector, which wasn´t part of the received samples.Thus, we were unable to further investigate the issue and the root cause remains unknown.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
|
| |
|