• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION HOMECHOICE CLARIA; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION HOMECHOICE CLARIA; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C6M40
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Dyspnea (1816)
Date of Event 04/03/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported an automated peritoneal dialysis patient experienced shortness of breath during a fill.The patient was connected to the homechoice claria device at the time of the event.The patient requested to end therapy and wanted to drain.Renal therapy services (rts) assisted the patient with ending therapy and a manual drain.The fill volume was 1970 ml.Rts initiated a swap of the device and discussed the use of continuous ambulatory peritoneal dialysis.There was no report of a medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
The device was not received for evaluation; therefore, a device analysis could not be completed; however, the sharesource log files associated with homechoice claria were reviewed.A review of the most recent therapy showed that the patient¿s average uf for the seven most recently completed therapies was approximately 650 ml and that the patient's programmed total uf (250 ml) was set too low for the most recent therapies.There was no device malfunction identified that could have caused or contributed to the reported issue.Based on the device log analysis, the most probable cause was determined to be the programmed total uf being set too low.Per the claria clinician guide, the product labelling warns users and clinicians that this may lead to accumulation of peritoneal fluid during the course of therapy, because of the undrained uf, which may cause symptoms in some patients.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE CLARIA
Common Device Name
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park d
singapore 73875 0
SN   738750
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key19183019
Report Number1416980-2024-01898
Device Sequence Number12540057
Product Code FKX
UDI-Device Identifier00085412676463
UDI-Public(01)00085412676463
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K201867
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number5C6M40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/03/2024
Supplement Date Received by Manufacturer05/31/2024
Initial Report FDA Received Date04/25/2024
Supplement Report FDA Received Date06/05/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-