The device was not received for evaluation; therefore, a device analysis could not be completed; however, the sharesource log files associated with homechoice claria were reviewed.A review of the most recent therapy showed that the patient¿s average uf for the seven most recently completed therapies was approximately 650 ml and that the patient's programmed total uf (250 ml) was set too low for the most recent therapies.There was no device malfunction identified that could have caused or contributed to the reported issue.Based on the device log analysis, the most probable cause was determined to be the programmed total uf being set too low.Per the claria clinician guide, the product labelling warns users and clinicians that this may lead to accumulation of peritoneal fluid during the course of therapy, because of the undrained uf, which may cause symptoms in some patients.Should additional relevant information become available, a supplemental report will be submitted.
|