| Model Number |
DISSECTOR, 4.0MM X 13CM |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a knee arthroscopy there was metal abrasion with the two shaver blades.There was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint allegation is confirmed.One unpackaged ar-8400ds serial/batch number 15150343 was received for evaluation.Functional testing was not performed due to damage to the device.Visual inspection found scratch lines inside and outside the outer tube, consistent with a site of metal debris production at the cutting head.The most likely cause can be attributed misuse due to prying/leveraging the device during insertion.
|
| |
|