Catalog Number 0603870
Medical Device Problem Codes Fluid/Blood Leak (1250); Fracture (1260); Difficult to Remove (1528); Material Separation (1562); Migration (4003)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 03/22/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that one month one day post a port placement, the port allegedly leaked.It was further reported that the catheter was allegedly found to be in the heart plane upon x-ray imaging.It was also reported that the catheter was allegedly unable to removed either through jugular or femoral veins and upon ct examination the catheter was allegedly found in the right and the left pulmonary artery.Reportedly, the port was removed and only the minor part of the catheter is fixed by the sleeve and the remaining part was allegedly torn away and migrated.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the m.R.I.Low-profile implantable port, groshong single-lumen, 7f products that are cleared in the us.The pro code and 510k number for the m.R.I.Low-profile implantable port, groshong single-lumen, 7f products is identified in d2 and g4.H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
 
Additional Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the m.R.I.Low-profile implantable port, groshong single-lumen, 7f products that are cleared in the us.The pro code and 510k number for the m.R.I.Low-profile implantable port, groshong single-lumen, 7f products is identified in d2 and g4.Manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported fluid leak, catheter fracture, material separation, migration and difficulty in removal issues as no objective evidence was provided for review.A definitive root cause could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that one month and one day post a port placement, the port allegedly leaked.It was further reported that the catheter was allegedly found to be in the heart plane upon x-ray imaging.It was also reported that the catheter was allegedly unable to be removed either through jugular or femoral veins and upon computed tomographic examination, the catheter was allegedly found in the right and the left pulmonary artery.Reportedly, the port was removed and only the minor part of the catheter was fixed by the sleeve and the remaining part was allegedly torn away and migrated.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported fluid leak, catheter fracture, material separation, migration and difficulty in removal issues as no objective evidence was provided for review.Furthermore, the clinical conditions alleged in the complaint cannot be confirmed.A definitive root cause could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.(expiration date: 07/2026), g3, h6(method).D4 (medical device catalog number, medical device lot number, unique identifier (udi) #).The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that one month and one day post a port placement, the port allegedly leaked.It was further reported that the catheter was allegedly found to be in the heart plane upon x-ray imaging.It was also reported that the catheter was allegedly unable to be removed either through jugular or femoral veins and upon computed tomographic examination, the catheter was allegedly found in the right and the left pulmonary artery.Reportedly, the port was removed and only the minor part of the catheter was fixed by the sleeve and the remaining part was allegedly torn away and migrated.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.R.I. LOW-PROFILE IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 7F
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key19183698
Report Number3006260740-2024-01922
Device Sequence Number15240641
Product Code LJT
UDI-Device Identifier00801741025839
UDI-Public(01)00801741025839
Combination Product (Y/N)N
Initial Reporter CountryRS
PMA/510(K) Number
K924250
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0603870
Device Lot NumberREFY2412
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/03/2024
Supplement Date Received by Manufacturer05/13/2024
07/04/2024
Initial Report FDA Received Date04/25/2024
Supplement Report FDA Received Date05/27/2024
07/11/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-