| Model Number |
TD-TB400 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that thunderbeat had problems with slow cutting sealing poorly and generator giving constant error messages due to the thunder beat transducer cable in a laparoscopic general surgical procedure (the exact procedure is not known).The procedure was successful, however the procedure took longer due to constant error messages.Follow up information received from the customer indicated that there were 62 error messages in one day predominantly u512 and u504.The cable was reconnected, and the generator restarted.In which 62 error messages occurred that interrupted and delayed the procedure.It was difficult to estimate the procedural delay in minutes, but it took a little over half an hour and at least 30 seconds per error message.The procedure was completed with the same device, and patient was unharmed.Reportedly, loud error messages appeared on the device, ¿causing the user to keep looking at the device to see what the error message and took someone out of their concentration.¿ no pre-existing conditions were known.No reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was returned to olympus for inspection, and there were no device abnormalities observed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been less than 1 year since the subject device was manufactured.Based on the results of the investigation, no device abnormalities were observed that could have led to the reported event.Therefore, the root cause could not be determined.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿chapter 8 ¿ troubleshooting: do not use the transducer on the patient if any irregularity is observed with the equipment.Otherwise, the system may not work properly and serious injury to the surgeon, surgical staff and/or the patient may result.¿ this supplemental report includes a correction to d9 and g2 from the initial medwatch.An update has been made to h3.Also, information has been made to h4.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|