| Model Number |
VICM512.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
03/12/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that a 12.6mm vicm512.6 implantable collamer lens of -10.50 diopter was implanted into the patient's left eye (os) on (b)(6) 2024.On (b)(6) 2024, the lens was exchanged for a same size, different power lens due to refractive surprise.Cause of the event is unknown.Status of the eye is "correct." the problem was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: work order search: no similar complaints within associated lots were found.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product evaluation - lens was returned dry in a microcentrifuge vial.Visual inspection found a haptic broken and stretched out lens.Dimensional and functional inspection found lens to be within specification.Claim# (b)(4).
|
| |
|