| Model Number |
701118150 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A reporter indicated "came away in the patient mid procedure after firing the needle.We manage to remove the core and then used another needle to take more samples.There was no harm to the patient.".
|
| |
|
Additional Manufacturer Narrative
|
|
The sample is indicated as returned to argon for investigation.A follow-up report will be submitted upon investigation completion.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the manufacturing and inspection records for this lot was conducted.No deviations or non-conformances were found.One opened sample was returned from the customer for review.A visual inspection was performed on the returned product.It was found that the plunger came off, inhibiting the device to function as intended.The complaint was confirmed.A sustaining engineering project has been initiated to investigate the issue in detail and identify the root cause and implement the corrective action to address the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi related data quality updates only.
|
| |
|