• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARGON MEDICAL DEVICES SUPERCORE BIOPSY INSTRUMENT 18GA X 15CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES SUPERCORE BIOPSY INSTRUMENT 18GA X 15CM Back to Search Results
Model Number 701118150
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/26/2024
Type of Reportable Event Malfunction
Event or Problem Description
A reporter indicated "came away in the patient mid procedure after firing the needle.We manage to remove the core and then used another needle to take more samples.There was no harm to the patient.".
 
Additional Manufacturer Narrative
The sample is indicated as returned to argon for investigation.A follow-up report will be submitted upon investigation completion.
 
Additional Manufacturer Narrative
A review of the manufacturing and inspection records for this lot was conducted.No deviations or non-conformances were found.One opened sample was returned from the customer for review.A visual inspection was performed on the returned product.It was found that the plunger came off, inhibiting the device to function as intended.The complaint was confirmed.A sustaining engineering project has been initiated to investigate the issue in detail and identify the root cause and implement the corrective action to address the issue.
 
Additional Manufacturer Narrative
Udi related data quality updates only.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERCORE BIOPSY INSTRUMENT 18GA X 15CM
Common Device Name
SUPERCORE BIOPSY
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
kelly mayer
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key19185954
Report Number0001625425-2024-00996
Device Sequence Number7147553
Product Code KNW
UDI-Device Identifier00886333005888
UDI-Public(01)00886333005888(17)281113(10)11506806
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K974814
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number701118150
Device Lot Number11506806
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/26/2024
Supplement Date Received by Manufacturer06/03/2024
08/31/2024
Initial Report FDA Received Date04/25/2024
Supplement Report FDA Received Date07/02/2024
08/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-