MAUDE Adverse Event Report: EXACTECH, INC. NV GXL LINER LIPPED 36MM ID GROUP 2; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. NV GXL LINER LIPPED 36MM ID GROUP 2; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 132-36-52
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Inflammation (1932); Pain (1994); Synovitis (2094); Osteolysis (2377)
Date of Event 02/06/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via legal documentation that a patient had a left total hip replacement procedure on (b)(6) 2018 and then was revised on (b)(6) 2024.Patient required revision surgery for polyethylene prosthesis wear.The mostly likely cause for the revision reported due to prothesis wear is a combination of risk factors including use error, implant positioning, implant size selection, and patient factors.However, this cannot be conclusively determined with the information provided.
 
Additional Manufacturer Narrative
D10: concomitant products: 4767729 164-01-14 - element-stem, collarless w/ha, std offset, sz 14.5456749 186-01-52 - integrip cc, cluster 52mm, g2.5471927 170-36-00 - biolox delta femoral head 36mm od, +0mm.H3: the revision reported was likely the result of prosthesis wear of the tibial insert.The cause of prosthesis wear is generally a combination of risk factors including use error, implant positioning, implant size selection, and patient factors.However, this cannot be conclusively determined with the information provided.
 
Additional Manufacturer Narrative
1038671-2024-00987 b1: corrected.B2: corrected.H6: corrected the following: health effect - impact code, type of investigation.Manufacturer narrative updated: based on the available information, the patient involved in the incident meets the following risk criteria for early prosthesis wear and/or osteolysis: combination of largest available femoral head and/or thinnest available acetabular liner was used.The most likely cause for the reported revision due to prosthesis wear and osteolysis is a combination of the risk factors such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed as the devices were not returned for evaluation, and images and radiographs were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LINER LIPPED 36MM ID GROUP 2
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
matt collins
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key19186996
Report Number1038671-2024-00987
Device Sequence Number15306857
Product Code JDI
UDI-Device Identifier10885862207036
UDI-Public10885862207036
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K121392
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 09/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/17/2023
Device Catalogue Number132-36-52
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2024
Supplement Date Received by Manufacturer08/14/2024
Initial Report FDA Received Date04/25/2024
Supplement Report FDA Received Date09/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/19/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1732-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-