| Catalog Number |
IAP-0500I |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported " the iabp doesn't work, low battery alarm".Additionally it is stated that this issue happened prior to use.No patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The reported complaint of "iabp doesn't work, low battery alarm" is not able to be confirmed.No part was returned to teleflex chelmsford for investigation.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.Iabp users and servicers must follow the operating instructions manual for recommendations on usage, charging, maintenance and storage of the batteries.If battery maintenance is not performed per the iabp operating instructions manual, the battery may not provide the expected minimum run time of operating power.
|
| |
|
Event or Problem Description
|
|
It was reported " the iabp doesn't work, low battery alarm".Additionally it is stated that this issue happened prior to use.No patient involvement reported.
|
| |
|