| Catalog Number |
LXMC16 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Inflammation (1932); Vomiting (2144); Insufficient Information (4580)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that it was seen in an x-ray that the linx device was discontinuous.Device is still implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 4/26/2024 b3: only event year known: 2024 d6a: exact date is unk.Assumed first month of the year and first day of the month.Lot number was received and dhr is pending review.When the review is completed, a supplemental medwatch will be sent with a summary of the evaluation.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 5/21/2024.Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.A manufacturing record evaluation was performed for the finished device 16424 number, and no non-conformances related to the malfunction were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.However, if the product is received at a later date, the investigation will be updated as applicable.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 7/1/2024 correction h12: d6a.Additional information was requested, and the following was obtained: what was the exact date of implant? ¿ (b)(6) 2017.What symptoms lead to the discovery of the discontinuous device? when did they begin? none ¿ was found incidentally on chest xray (b)(6) 2023 what was the date of the imaging which showed the discontinuous linx? if available, please share a copy of this imaging.Please send to: productcompliant1@its.Jnj.Com ¿ chest xray dated (b)(6) 2023, same image as (b)(6) 2023 was the device initially effective in controlling reflux? yes but by 2019 egd by an outside gi noted la-grade b esophagitis were any events associated with the onset of symptoms (vomiting, retching, trauma, surgery)? difficult to tell, patient has cyclic vomiting syndrome did the patient undergo an mri since device implant? if so, when was the mri and what strength? mri brain, abdomen (b)(6) 2018 1.5t.Mri brain (b)(6) 2018 1.5 t, mri brain (b)(6) 2018 1.5 t, (b)(6) 2020 mri lumbar spine 1.5t did the patient have any other surgeries in the area? no was any additional imaging performed since device implant? does the device appear to be in a continuous annular state in these images? we are interested in establishing a window when the device may have become discontinuous.Please share any additional images.Ct abdomen 2020 appears normal as do the cxr in 2020 and 2019 what is the management plan? is device removal scheduled? is a replacement linx or fundoplication planned? patient elects to be monitored conservatively.She has multiple medical problems and was only curious/concerned if she should have it removed/replaced but is not seeking surgery if not necessary when and if the explanation takes place can we ask that the procedure gets video recorded and the video shared? will need to ask administration if we get to this point photo analysis: x-ray images of the device in vivo was reviewed by a medical safety officer.As per medical safety officer: "the first image demonstrates a discontinuous device.The images labeled 2019 and 2020 appear to demonstrate intact linx devices.".
|
| |
|
Search Alerts/Recalls
|