| Model Number |
PEEK SWIVELOCK SP, 5.5X24.5MM |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint device was not received for evaluation.Based off the information provided, the most likely cause for the reported failure can be attributed to improper bone prep; misaligned insertion; prying/leveraging the device during insertion.
|
| |
|
Event or Problem Description
|
|
The surgeon tried to implant 2 x 5.5 self punching swivelocks in the manner he would normally use them for a lateral row in an open rotator cuff repair and the tips bent.
|
| |
|
Search Alerts/Recalls
|