| Model Number |
N/A |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that tech noticed a hole in the packaging when the implant was opened.Another implant was used to complete the surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected}.Visual evaluation of the provided photos and returned product found puncture marks in the inner foil lid.Sterility is considered not breached as no damage to the outer cavity was reported.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.Upon reassessment of the reported event, it was determined to be not reportable as sterility was not breached.The initial report should be voided.
|
| |
|
Event or Problem Description
|
|
No further information at the time of this report.
|
| |
|