Catalog Number 02-012-35-3011
Medical Device Problem Codes Mechanical Problem (1384); Unstable (1667); Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Loss of Range of Motion (2032); Joint Laxity (4526); Implant Pain (4561); Swelling/ Edema (4577)
Date of Event 03/14/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: concomitant devices: 2584040 02-012-45-3020 - lgc tibial fit tray cem sz 3f / 2t.2830911 200-02-35 - three peg patella 35mm.2857570 02-010-01-0230 - logic femoral ps cem left sz 3.9y088 203-96-03 - (11-2216) saw blade new stryker (.050).9y090 203-96-01 - (11-2222) saw blade new stryk (.050).H3: the revision reported was likely the result of prosthesis wear of the tibial insert.The cause of prosthesis wear is generally a combination of risk factors including, use error, implant positioning, implant size selection, and patient factors.However, this cannot be conclusively determined with the information provided.
 
Event or Problem Description
As reported via legal documentation, a patient had a left total knee replacement procedure on (b)(6) 2013 and then was revised on (b)(6) 2023.The patient required revision surgery for issues including but not limited to significant pain in her left knee, along with swelling, instability, limited range of motion, and loss of mobility.The mostly likely cause for the revision reported due to prosthesis wear is a combination of risk factors including use error, implant positioning, implant size selection, and patient factors.However, this cannot be conclusively determined with the information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA PS MOD INSRT SZ 3 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
matt collins
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key19191874
Report Number1038671-2024-00990
Device Sequence Number8360980
Product Code JWH
UDI-Device Identifier10885862001818
UDI-Public10885862001818
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 11/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/19/2021
Device Catalogue Number02-012-35-3011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2024
Supplement Date Received by Manufacturer11/08/2024
Initial Report FDA Received Date04/26/2024
Supplement Report FDA Received Date11/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/22/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0021-2022
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-