| Model Number |
N/A |
| Medical Device Problem Codes |
Incorrect Measurement (1383); Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the posterior referencing sizer seems a bit too loose when used on the femur and isn¿t as precise as it should be.Another device was used to complete the surgery.There was no impact on the patient.The surgical technique was utilized.There was no surgical delay.Attempts have been made and all available information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).G2 : foreign country : canada.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g1; g3; g6; h1; h2; h3; h6 evaluation of the returned device exhibits signs of use and the two components rotate a little at each setting.The device is conforming.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.The device specifications were found to be conforming per the functional check and hence no issue was seen with the device.A field communication was released notifying users of appropriate use of the device according to the surgical technique.Upon receiving additional information of the reported event, it was determined to be not reportable as there was no problem found with the device.The initial report should be voided.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|