The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that the user inserted a coude tip catheter into patient.Inflated balloon with 10ml ns.Patient had felt a "pop".When the procedure was complete, the user pulled catheter and the silicone of where balloon was hanging off catheter tubing.Patient did not have any pain with same and had inconvenience.Per additional information received on 08apr2024, it was reported that after removing the catheter after it had popped, the silicone piece where the balloon was had obviously broken since it had burst while in the patient.
|
|
It was reported that the user inserted a coude tip catheter into patient.Inflated balloon with 10ml ns.Patient had felt a pop.When the procedure was complete, the user pulled catheter and the silicone of where balloon was hanging off catheter tubing.Patient did not have any pain with same and had inconvenience.Per additional information received on 08apr2024, it was reported that after removing the catheter after it had popped, the silicone piece where the balloon was had obviously broken since it had burst while in the patient.
|
|
The reported event was confirmed cause unknown.Visual evaluation of the returned photo sample noted one opened (with original packaging), used foley catheter.Visual inspection of the photo sample noted the balloon was rupture.It was also received 1 foley catheter with original packaging.Verified material number 0102l12 and manufacturing lot number nghu2593.Visual inspection noted the balloon had rupture.Further inspection noted the balloon had no missing pieces.The exact root cause could not be identified.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be high modulus latex.However, there was insufficient information to confirm this potential root cause.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: caution: this product contains natural rubber latex which may cause allergic reactions.Sterile unless package is opened or damaged.Warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage latex and may cause the balloon to burst.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Single patient use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip tip syringe.Do not use needle.Recommended inflation capacities 3cc balloon: use 5ml sterile water 5cc balloon: use 10ml sterile water 15cc balloon: use 20ml sterile water 20cc balloon: use 25ml sterile water 30cc balloon: use 35ml sterile water 40cc balloon: use 45ml sterile water 75cc balloon: use 80ml sterile water do not exceed recommended capacities.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe stick in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.Udi format updated with available product information per gudid as requested.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|