Catalog Number DR8042
Medical Device Problem Codes Tear, Rip or Hole in Device Packaging (2385); Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, a photo was provided for review.The investigation of the reported event is currently underway.H10: d4 (expiration date: 11/2026).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that prior to an angioplasty procedure, the package of the pta balloon was allegedly noted bent in the middle.It was further reported that breach in the sterility was allegedly noted.There was no patient contact.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one dorado inner carton was returned without the outer package.The inner carton was noted to be in a sealed condition.During visual analysis the inner carton package was received in a folded condition which caused the kinking to the device and protective tube in which no evidence of package issue could not be noted, no breach of sterility loss could be observed.No other anomalies were noted.Also, one photo was received for review.In the photo outer package of product and label details showed which matched with complaint record.Further in outer package showed in photo has no severe kink nor dent mark noted and no evidence of breach of sterility could be observed.No other anomalies were observed.Based on the submitted photo and received sample no evidence of packaging problem could be observed, due to the packaging condition and no breach of sterility noted to the returned inner carton.Hence the investigation was inconclusive for the reported packaging problem and unconfirmed for the breach of sterility.A definitive root cause for the reported packaging problem and breach of sterility issue could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 11/2026).H11: e3, h6 (method, result, conclusion).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that prior to an angioplasty procedure, the package of the pta balloon was allegedly noted bent in the middle.It was further reported that breach in the sterility was allegedly noted.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens 75751
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key19192880
Report Number2020394-2024-00573
Device Sequence Number8361525
Product Code LIT
UDI-Device Identifier00801741058264
UDI-Public(01)00801741058264
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K072283
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberDR8042
Device Lot Number93YH0109
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/03/2024
Supplement Date Received by Manufacturer06/04/2024
Initial Report FDA Received Date04/26/2024
Supplement Report FDA Received Date06/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-