| Catalog Number |
106696 |
| Medical Device Problem Codes |
Break (1069); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the "pressure bypass valve" of a prismaflex m60 set was broken and leaking during priming.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: the device was received for evaluation.During visual inspection, it was observed that the cone of the pre-blood pump male luer lock was broken.This component is provided by a baxter supplier.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the luer connector damage was determined to be related to supplier product design and manufacturing.A corrective action preventive action record and a change control were previously opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|