Model Number 4.75MM BC KNOTLESS SWIVELOCK
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/29/2021
Type of Reportable Event Malfunction
Event or Problem Description
On 12/29/2021, it was reported by a sales representative that an ar-2324kbcc knotless swivelock pulled out of implantation site.This was discovered during a metacarpals joint arthroplasty procedure on (b)(6) 2021.When surgeon attempted to pull the shuttling stitch that pulls the repair stitch through the anchor, it got stuck.When surgeon applied a bit more of force, the anchor pulled out.Case was completed successfully without further issues using a regular 5.5mm swivelock and inserter in the same previously prepared bone hole.
 
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint device was not received for evaluation.Based off the information provided, the most likely cause for the reported failure can be attributed to user error of the device due to excessive force used during suture passing/tightening /tensioning.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.75MM BC KNOTLESS SWIVELOCK
Common Device Name
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19193878
Report Number1220246-2024-02475
Device Sequence Number20922965
Product Code MAI
UDI-Device Identifier10888867351285
UDI-Public10888867351285
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K193503
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2025
Device Model Number4.75MM BC KNOTLESS SWIVELOCK
Device Catalogue NumberAR-2324KBCC
Device Lot Number13675704
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/29/2021
Initial Report FDA Received Date04/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-