• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. KNEE COMPONENT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. KNEE COMPONENT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-44-5009
Medical Device Problem Codes Unstable (1667); Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Synovitis (2094); Discomfort (2330); Joint Laxity (4526); Swelling/ Edema (4577)
Date of Event 03/22/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via legal documentation that a patient had a right total knee replacement procedure on (b)(6) 2015 and then was revised on (b)(6) 2023.Patient required revision surgery for issues including but not limited to pain, swelling, instability, and polyethylene wear.The mostly likely cause for the revision reported due to prothesis wear is a combination of risk factors including use error, implant positioning, implant size selection, and patient factors.However, this cannot be conclusively determined with the information provided.
 
Additional Manufacturer Narrative
The revision reported was likely the result of prosthesis wear of the tibial insert.The cause of prosthesis wear is generally a combination of risk factors including, use error, implant positioning, implant size selection, and patient factors.However, this cannot be conclusively determined with the information provided.
 
Additional Manufacturer Narrative
H6: corrected the following: health effect clinical code, medical device problem code, type of investigation.
 
Additional Manufacturer Narrative
A1-3: added patient age, dob and gender.B1-2: corrected product problem and corrected outcomes to hospitalization and required intervention.D4: corrected catalogue, serial, and udi numbers.Corrected expiration date.D8: corrected "no" to "used by 3rd party?" g: corrected reporter contact and email.G8: corrected 501k.H6: added health impact clinical codes 2330-2094.D10: concomitant devices: (b)(6) 02-010-01-0350 - logic femoral ps cem right sz 5.(b)(6) 02-012-45-5050 - lgc tibial fit tray cem sz 5f / 5t.(b)(6) 200-02-35 - three peg patella 35mm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KNEE COMPONENT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key19194934
Report Number1038671-2024-00995
Device Sequence Number20950788
Product Code JWH
UDI-Device Identifier10885862173713
UDI-Public10885862173713
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K110547
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/07/2023
Device Catalogue Number02-012-44-5009
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2024
Supplement Date Received by Manufacturer09/24/2024
Not provided
Initial Report FDA Received Date04/26/2024
Supplement Report FDA Received Date09/27/2024
11/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age58 YR
Patient SexMale
-
-