• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 20MM CENTRAL POST, MODULAR; SHOULDR PROSTH, REVERSE CONFIG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 20MM CENTRAL POST, MODULAR; SHOULDR PROSTH, REVERSE CONFIG Back to Search Results
Model Number 20MM CENTRAL POST, MODULAR
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint device was not received for evaluation.Based off the information provided, the most likely cause for the reported failure can be attributed to improper bone prep and/or misaligned insertion.
 
Event or Problem Description
On 10/28/2021, it was reported by a sales representative via phone that an ar-9105-03 glenoid didn't fit and the ar-9561-20p central post would not engage.This was discovered during use in a reverse arthroplasty on (b)(6) 2021.The case was completed by using a new ar-9105-03 and ar-9561-20p.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
20MM CENTRAL POST, MODULAR
Common Device Name
SHOULDR PROSTH, REVERSE CONFIG
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19195442
Report Number1220246-2024-02485
Device Sequence Number8474422
Product Code PHX
UDI-Device Identifier00888867296510
UDI-Public00888867296510
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K173900
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2025
Device Model Number20MM CENTRAL POST, MODULAR
Device Catalogue NumberAR-9561-20P
Device Lot Number8961
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/28/2021
Initial Report FDA Received Date04/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-