| Model Number |
SR-0940-CS |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Deformation (2976); Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product associated with this complaint is pending return.There was no patient harm reported, however, intervention was taken to prevent injury and therefore this complaint will be reported out of abundance of caution.All reasonably available information has been provided by the company at the time of submission of this report.The fields that are blank are not an omission and indicate that the information is either not applicable or currently unavailable.If additional information is received, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that during a left transcarotid artery revascularization (tcar) procedure, while removing the enroute transcarotid stent system (tss), the tss caught on the stent and pulled the distal portion of the stent into the proximal segment.The stent was excised, and an unplanned additional stent was placed.The procedure was completed successfully with no further complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The inspection of the returned stent delivery system (sds) did not identify any non-conformances; however, the stent was not returned for analysis.The customer reported complaint could not be confirmed.The exact cause of the reported event could not be conclusively determined during the analysis.
|
| |
|
Event or Problem Description
|
|
This supplemental mdr is being submitted to update the h11 device investigation results for the initial report submitted on 04/26/2024.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information can be found in the following fields: b4: date of this report.G3: date received by manufacturer.G6: type of report.H2: if follow up, what type? h3: device evaluated by manufacturer? h6: type of investigation.H11: additional manufacturer narrative.The stent delivery system (sds) and deployed stent were returned for analysis.The complaint reported by the customer regarding difficult removal and unintended movement of the delivery system could not be confirmed.The dimensional analysis indicated the device to be within specifications.The stent was found to be fractured upon return.The customer did not report a fractured stent, and the cause of the fracture could not be definitively determined.It was reported that during the procedure, the stent was pulled in on itself during withdrawal from the delivery system.The procedural factors and handling associated with this may have contributed to the observed fracture.The product analysis does not suggest that the observed damage is related to the manufacturing process.
|
| |
|
Event or Problem Description
|
|
This supplemental mdr is being submitted to update the h11 device investigation results for the initial report submitted on 04/26/2024.
|
| |
|