• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. TRULIANT TIB IMP CRC INSERT SZ 3, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. TRULIANT TIB IMP CRC INSERT SZ 3, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-022-51-3009
Medical Device Problem Codes Unstable (1667); Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Inflammation (1932); Pain (1994); Scar Tissue (2060); Numbness (2415); Joint Laxity (4526); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 06/21/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: concomitant devices: 03009021137 a10012 - gps implant kit v2 west, cynthia.5681453 02-020-14-0330 - truliant cr por fem cr por right sz 3 west, cynthia.6737458 201-78-98 - 2 pk, schanz pin, 4mm x 130mm west, cynthia.6841260 02-022-55-3020 - truliant por tib tray size 3f/2t west, cynthia.6907733 201-78-15 - holding pin mini sharp point 4 pk west, cynthia.S265322 521-78-32 - threaded pin size 3.0 collarless 2pk west, cynthia.S265534 521-78-32 - threaded pin size 3.0 collarless 2pk west, cynthia.S265576 521-78-23 - threaded pin size 2.3 collared 2pk west, cynthia.H3: the revision reported was likely the result of prosthesis wear of the tibial insert.The cause of prosthesis wear is generally a combination of risk factors including, use error, implant positioning, implant size selection, and patient factors.However, this cannot be conclusively determined with the information provided.
 
Event or Problem Description
It was reported via legal documentation that a patient had a right knee replacement procedure on (b)(6) 2021 and then was revised on (b)(6) 2023.Patient required revision surgery for issues including but not limited to pain, swelling, scarring, numbness, inflammation, knee popping and clicking, instability, difficulty with activities of daily living.The mostly likely cause for the revision reported due to prosthesis wear is a combination of risk factors including use error, implant positioning, implant size selection, and patient factors.However, this cannot be conclusively determined with the information provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were corrected: b1, b2, h6.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and instability or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
Concomitant devices: (b)(6) a10012 - gps implant kit v2; (b)(6) 02-020-14-0330 - truliant cr por fem cr por right sz 3; (b)(6) 201-78-98 - 2 pk, schanz pin, 4mm x 130mm; (b)(6) 02-022-55-3020 - truliant por tib tray size 3f/2t; (b)(6) 201-78-15 - holding pin mini sharp point 4 p; (b)(6) 521-78-32 - threaded pin size 3.0 collarless 2pk; (b)(6) 521-78-32 - threaded pin size 3.0 collarless 2pk; (b)(6) 521-78-23 - threaded pin size 2.3 collared 2pk.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and instability or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT TIB IMP CRC INSERT SZ 3, 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key19197057
Report Number1038671-2024-00998
Device Sequence Number12672673
Product Code JWH
UDI-Device Identifier10885862307637
UDI-Public10885862307637
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K152170
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number02-022-51-3009
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/05/2024
Supplement Date Received by Manufacturer11/22/2024
11/22/2024
Initial Report FDA Received Date04/26/2024
Supplement Report FDA Received Date11/26/2024
11/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0023-2022
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-