| Model Number |
2110 |
| Medical Device Problem Codes |
Device Alarm System (1012); Blocked Connection (2888)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device had an "occlusion alarm ".There was patient involvement and no patient harm/adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received for investigation.The device was visually inspected and functionally tested.The reported problem was not duplicated during testing.However, the problem was identified during ehl inspection.The root cause is a faulty dso sensor.This will be replaced.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Reporter phone: (b)(6).
|
| |
|