Model Number 2110
Device Problems Device Alarm System (1012); Blocked Connection (2888)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/24/2024
Event Type  malfunction  
Event Description
It was reported that the device had an "occlusion alarm ".There was patient involvement and no patient harm/adverse event reported.
 
Manufacturer Narrative
One device was received for investigation.The device was visually inspected and functionally tested.The reported problem was not duplicated during testing.However, the problem was identified during ehl inspection.The root cause is a faulty dso sensor.This will be replaced.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Reporter phone: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD SOLIS HPCA AMBULATORY INFUSION PUMP
Type of Device
PUMP, INFUSION, PCA
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key19198057
MDR Text Key341371347
Report Number3012307300-2024-03123
Device Sequence Number1
Product Code MEA
UDI-Device Identifier10610586038822
UDI-Public10610586038822
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K130394
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2110
Device Catalogue Number21-2112-0300-02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/28/2024
Initial Date FDA Received04/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/06/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-