| Catalog Number |
MPX5302-C |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Material#: mpx5302-c.Batch#: unknown.It was reported by customer that the side port of the extension set oozes blood while flushing.I take it that she is hooking a vacutainer to the side, proximal port and drawing blood and then flushing the distal port with saline.I don't know if this process is damaging the internal valve on the prox.Side port and causing the issue.Verbatim: rcc received a complaint via email.Email(s) attached.Good morning christine.One of our teammates has reported that the side port of the extension set oozes blood while flushing.I take it that she is hooking a vacutainer to the side, proximal port and drawing blood and then flushing the distal port with saline.I don't know if this process is damaging the internal valve on the prox.Side port and causing the issue.Do you have any guidance on this issue or can put me in contact with the rep who handles this device? thank you! md max plus ext set mpx5302-c.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.The actual date of event is unknown.The date received by manufacturer was entered into the date of event field.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer reported leakage from the port, and returned videos of the issue.The videos verify leaking from the blue piston on the y-site of mpx5302-c.The physical sample was unable to be returned, and no device investigation was able to be performed.A device history record review could not be performed on material mpx5302-c because the lot number is unknown.
|
| |
|
Event or Problem Description
|
|
No additional info.
|
| |
|
Search Alerts/Recalls
|