| Catalog Number |
11522558 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
| |
|
Event or Problem Description
|
|
Material#: 11522558.Batch number#: 22095036.It was reported by customer that issue has been reported regarding potential leaking.Verbatim#: rcc received a complaint via email.Our customer has lodged a complaint regarding item#: 11522558, lot#: 22095036.The complaint is as follows: please note that an issue has been reported regarding potential leaking of l 135161, c # (b)(4) iv set pump w/ball 3 port with lot number (10) 22095036.Value analysis has requested this item and lot number be pulled from use and returned to concordance for return to the manufacturer.Please provide rga for the return of 601 ea of this item.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported by customer that issue has been reported regarding potential leaking.No product or photo was returned by the customer.The customer complaint of leakage could not be verified due to the product not being returned for failure investigation.Due to no sample being received, an investigation could not be performed and a root cause could not be determined.A device history record review for model 11522558 lot number 22095036 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 01sep2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
No additional information was provided.
|
| |
|
Event or Problem Description
|
|
No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported by customer that issue has been reported regarding potential leaking.Three unused samples were received for quality investigation.Visual inspection of the packages and infusion sets were conducted.There were indications of damages or abnormalities on the infusion sets.The infusion sets were then primed with water to determine if there were any issues during priming.No issues were discovered during priming.A simulated infusion was then conducted with the infusion sets with a flow rate of 125 ml/hr for 1 hour per set.There were not issues found during the simulated infusions.The customer complaint of leakage could not be confirmed.A root cause was not established because the reported failure could not be replicated.A device history record review for model 11522558 lot number 22095036 was performed.The search showed that a total of 11,523 units in 1 lot number was built on 01sep2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|