• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON AS LVP 20D 3SS CV BV; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON AS LVP 20D 3SS CV BV; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 11522558
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Event or Problem Description
Material#: 11522558.Batch number#: 22095036.It was reported by customer that issue has been reported regarding potential leaking.Verbatim#: rcc received a complaint via email.Our customer has lodged a complaint regarding item#: 11522558, lot#: 22095036.The complaint is as follows: please note that an issue has been reported regarding potential leaking of l 135161, c # (b)(4) iv set pump w/ball 3 port with lot number (10) 22095036.Value analysis has requested this item and lot number be pulled from use and returned to concordance for return to the manufacturer.Please provide rga for the return of 601 ea of this item.
 
Additional Manufacturer Narrative
It was reported by customer that issue has been reported regarding potential leaking.No product or photo was returned by the customer.The customer complaint of leakage could not be verified due to the product not being returned for failure investigation.Due to no sample being received, an investigation could not be performed and a root cause could not be determined.A device history record review for model 11522558 lot number 22095036 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 01sep2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
No additional information was provided.
 
Event or Problem Description
No additional information was provided.
 
Additional Manufacturer Narrative
It was reported by customer that issue has been reported regarding potential leaking.Three unused samples were received for quality investigation.Visual inspection of the packages and infusion sets were conducted.There were indications of damages or abnormalities on the infusion sets.The infusion sets were then primed with water to determine if there were any issues during priming.No issues were discovered during priming.A simulated infusion was then conducted with the infusion sets with a flow rate of 125 ml/hr for 1 hour per set.There were not issues found during the simulated infusions.The customer complaint of leakage could not be confirmed.A root cause was not established because the reported failure could not be replicated.A device history record review for model 11522558 lot number 22095036 was performed.The search showed that a total of 11,523 units in 1 lot number was built on 01sep2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS LVP 20D 3SS CV BV
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key19199069
Report Number9616066-2024-00652
Device Sequence Number7620047
Product Code FPA
UDI-Device Identifier10885403232305
UDI-Public(01)10885403232305(11)220829(17)250901(10)22095036
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K022209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2025
Device Catalogue Number11522558
Device Lot Number22095036
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2024
Supplement Date Received by Manufacturer07/08/2024
09/23/2024
Initial Report FDA Received Date04/28/2024
Supplement Report FDA Received Date07/15/2024
10/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-