• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CRITICAL CARE AB FLOW-E ANESTHESIA SYSTEM; GAS-MACHINE, ANESTHESIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CRITICAL CARE AB FLOW-E ANESTHESIA SYSTEM; GAS-MACHINE, ANESTHESIA Back to Search Results
Model Number 6887900
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/06/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the anesthesia system made a loud clicking noise and did not deliver breaths.Airway pressure alarms were generated.There was no patient harm.Manufacturer's reference number: (b)(4).
 
Additional Manufacturer Narrative
On-site investigation was performed by our field service engineer.The claimed issue was not reproduced during troubleshooting.Based on provided information after switching to manual ventilation and back to controlled mode, the issue did not reoccur and device was working correctly.Patient was connected back to anesthesia workstation and case was continue without interruption.Reported issue can be confirmed in provided device's logs.Evaluation of the logs showed that on the date of event system checkout (sco) was passed prior to the case.The leakage check was also performed.After case device also passed system checkout.The most probable cause of the high airway pressure alarm and clicking noise during case was the lung compliance was lower than expected causing the pressure regulated alarms and clicking noise was caused by safety valve being open due to activated high airway pressure alarm.Our conclusion is that changing the ventilation mode solved the reported failure.A correction of fields #d1 brand name, #d4 version or model#, #d4 unique identifier (udi) #, #h4 device manufacture date, #h6 adverse event problem and evaluation codes - health effect ¿ impact codes.D1 - brand name - previous brand name: flow-e, corrected brand name: flow-e anesthesia system.D4 - version or model # - previous version or model #.Flow-e.Corrected version or model #: 6887900.D4 ¿ unique identifier (udi) # - previous unique identifier (udi) #: missing, corrected unique identifier (udi) #: (b)(4).H4 - device manufacture date - previous manufacture date: 11/01/2021, corrected manufacture date: 10/21/2021.H6 - adverse event problem and evaluation codes - health effect ¿ impact codes - previous health effect ¿ impact codes: missing, corrected health effect ¿ impact codes: no health consequences or impact|||2199.
 
Event or Problem Description
Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOW-E ANESTHESIA SYSTEM
Common Device Name
GAS-MACHINE, ANESTHESIA
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
caroline kabbabe
roentgenvagen 2
solna 
MDR Report Key19199709
Report Number8010042-2024-00675
Device Sequence Number8034572
Product Code BSZ
UDI-Device Identifier07325710010457
UDI-Public(01)07325710010457(11)211021
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K191027
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6887900
Device Catalogue Number6887900
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2024
Supplement Date Received by Manufacturer06/27/2024
Initial Report FDA Received Date04/29/2024
Supplement Report FDA Received Date07/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/21/2021
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-