| Catalog Number |
515064 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.H3 other text : see h10 manufacture narrative.
|
| |
|
Event or Problem Description
|
|
Material #515064.Lot # 2308404.It was reported by customer that they noticed some black like residue all along the inner silicone that covers the spike.Verbatim: rcc received a complaint via email.Email(s) attached today while i was compounding, when i removed the green cap from the p20 protector ref515064 i noticed some black like residue all along the inner silicone that covers the spike.26-apr: fu sent.Customer response on april 26, 2024."the product was not used as the technician noticed this after unpacking and taking off the cap and getting ready to spike the vial.".
|
| |
|
Additional Manufacturer Narrative
|
|
Pr (b)(4) follow up mdr for device evaluation: both the physical samples along with photos were provided to our quality team for investigation.Through visual inspection, a stain is observed on the protector sleeve.Further evaluation determined the stain is grease.A device history review was performed for reported lot 2308404, no deviations or non-conformances related to the reported issue were identified during the manufacturing process.Retained samples of the reported lot were used for additional evaluation.All products were inspected, no foreign matter or grease stains were observed on any of the devices.Product undergoes inspections throughout the manufacturing process to ensure the quality and functionality of the device, no issues related to this incident were identified.Machines undergo routine cleaning and maintenance according to procedure, all inspection records verify that all quality processes were carried out normally.While we could not identify a direct issue, it was determined the grease stain likely occurred during the molding process as a result of the product coming into contact with a specific part of the equipment that must be greased for operation.Manufacturing personnel have been notified of this incident.Complaints received for this device and reported condition will be monitored by our quality team for signs of emerging trends.
|
| |
|
Event or Problem Description
|
|
Material #515064, lot # 2308404.It was reported by customer that they noticed some black like residue all along the inner silicone that covers the spike.Verbatim: rcc received a complaint via email.Email(s) attached.Today while i was compounding, when i removed the green cap from the p20 protector ref515064 i noticed some black like residue all along the inner silicone that covers the spike.26-apr: fu sent.Customer response on (b)(6) 2024."the product was not used as the technician noticed this after unpacking and taking off the cap and getting ready to spike the vial.".
|
| |
|