| Catalog Number |
4252535-02 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
04/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number 400653090.The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: the needle broke in the patient angio sheared from the needle patient was fine but complaining of pain.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Photographs were sent for evaluation.The provided pictures was visually evaluated for damaged.It was noted the introcan safety needle was used and the needle punctured the catheter at the tip.The reported defect was not confirmed.All available information was forwarded to the supplier for further evaluation.A possible root cause is not considered to be during the manufacturing process.It should be noted that the affected sample is used, the most probable root cause is that wrong caused the breakage noted in the picture.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|