B. BRAUN MEDICAL INC. INTROCAN SAFETY®; CATHETER,INTRAVASCULAR,THE Back to Search Results
Catalog Number 4252535-02
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Pain (1994)
Date of Event 04/15/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number 400653090.The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: the needle broke in the patient angio sheared from the needle patient was fine but complaining of pain.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Photographs were sent for evaluation.The provided pictures was visually evaluated for damaged.It was noted the introcan safety needle was used and the needle punctured the catheter at the tip.The reported defect was not confirmed.All available information was forwarded to the supplier for further evaluation.A possible root cause is not considered to be during the manufacturing process.It should be noted that the affected sample is used, the most probable root cause is that wrong caused the breakage noted in the picture.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY®
Common Device Name
CATHETER,INTRAVASCULAR,THE
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key19200829
Report Number2523676-2024-00407
Device Sequence Number20815156
Product Code FOZ
UDI-Device Identifier04046963165925
UDI-Public(01)04046963165925
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K020785
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number4252535-02
Device Lot Number0061918660
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/29/2024
Supplement Report FDA Received Date08/28/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-