As reported, after implantation of an 8mm x 100mm smart radianz self-expanding stent (ses), when removing the delivery system from the patient over a 300cm vassallo floppy guidewire, it was noticed that the front end was torn off.The 10cm separated segment remained on the vassallo guidewire and was able to be removed without issue by pulling the guidewire.Additionally, there was some resistance when removing the delivery system due to resistance with the guidewire.The procedure continued by implanting a 7mm x 60mm smart vascular ses at the external iliac artery (eia).The right common iliac artery was then treated, and the procedure was completed.There were no reported injuries to the patient.This was during a procedure to treat two separate lesions, one in the left common iliac artery to external iliac artery (eia), and one in the right common iliac artery to superficial femoral artery (sfa).The eia lesion had a stenosis of 85% - 95% and was moderately calcified.The sfa lesion had a stenosis of 90%-100% and was moderately calcified.A 6f 90cm non-cordis guiding sheath was used for access from the left brachial artery to the left iliac artery.Various unknown guidewires were used and were able to pass.An unknown balloon catheter was used for pre-dilation.The smart radianz ses delivery system was stored and prepped per the instructions for use (ifu).The device will be returned for evaluation.
|
|
As reported, after implantation of an 8mm x 100mm smart radianz self-expanding stent (ses), when removing the delivery system from the patient over a 300cm vassallo floppy guidewire, it was noticed that the front end was torn off.The 10cm separated segment remained on the vassallo guidewire and was able to be removed without issue by pulling the guidewire.Additionally, there was some resistance when removing the delivery system due to resistance with the guidewire.The procedure continued by implanting a 7mm x 60mm smart vascular ses at the external iliac artery (eia).The right common iliac artery was then treated, and the procedure was completed.There were no reported injuries to the patient.This was during a procedure to treat two separate lesions, one in the left common iliac artery to external iliac artery (eia), and one in the right common iliac artery to superficial femoral artery (sfa).The eia lesion had a stenosis of 85% - 95% and was moderately calcified.The sfa lesion had a stenosis of 90%-100% and was moderately calcified.A 6f 90cm non-cordis guiding sheath was used for access from the left brachial artery to the left iliac artery.Various unknown guidewires were used and were able to pass.An unknown balloon catheter was used for pre-dilation.The smart radianz ses delivery system was stored and prepped per the instructions for use (ifu) and there were no difficulties flushing the device.The device will be returned for evaluation.
|
|
After further review of additional information received the following sections have been updated accordingly: b4, g3, g6, h1, h2, h3, h6, and h10.Complaint conclusion: as reported, after implantation of an 8mm x 100mm smart radianz self-expanding stent (ses), when removing the delivery system from the patient over a 300cm vassallo floppy guidewire, it was noticed that the front end was torn off.The 10cm separated segment remained on the vassallo guidewire and was able to be removed without issue by pulling the guidewire.Additionally, there was some resistance when removing the delivery system due to resistance with the guidewire.The procedure continued by implanting a 7mm x 60mm smart vascular ses at the external iliac artery (eia).The right common iliac artery was then treated, and the procedure was completed.There were no reported injuries to the patient.This was during a procedure to treat two separate lesions, one in the left common iliac artery to external iliac artery (eia), and one in the right common iliac artery to superficial femoral artery (sfa).The eia lesion had a stenosis of 85% - 95% and was moderately calcified.The sfa lesion had a stenosis of 90%-100% and was moderately calcified.A 6f 90cm non-cordis guiding sheath was used for access from the left brachial artery to the left iliac artery.Various unknown guidewires were used and were able to pass.An unknown balloon catheter was used for pre-dilation.The smart radianz ses delivery system was stored and prepped per the instructions for use (ifu) and there were no difficulties flushing the device.The device was returned for analysis.A non-sterile ¿smart radianz 8x100 190cm¿ was received for analysis inside of a plastic bag.The device was unpacked and was placed on a metallic tray to perform the product evaluation.The unit was inspected observing the following conditions: the stent was fully deployed and not returned for analysis.The distal tip is separated, and the missing piece was not returned for analysis.No other outstanding details were observed on the device.Functional analysis was performed by insertion/withdrawal testing to determine if resistance/friction or an obstruction between the wire lumen and the guidewire was noted.A lab sample guidewire of the proper size was inserted by the distal tip to be advanced all the way through the inner shaft of the device.The guidewire traveled within the lumen of the wire until it could be seen emerging from the exit port.Despite the distal tip separation, no resistance, friction, or obstruction was observed during the functional test.Sem analysis was not performed because the damages associated with the separation are visible with the magnification obtained with a vision system.The separated area was inspected observing evidence of elongations, and plastic deformations.The braid wire shows a reduction in diameter.The separated edges have an irregular path.The damages found are commonly associated with separations caused by material tensile overload.Therefore, it is assumed the device was induced to a tensile force that exceeded the material yield strength prior to the separation.No other anomalies were observed during the evaluation.The reported ¿guidewire lumen (inner shaft) resistance/friction¿ was not confirmed as the device passed functional analysis via insertion/withdrawal testing despite the missing distal portion of the device.The reported ¿catheter tip separated¿ was confirmed as the distal portion of the catheter tip was not returned with the product for analysis.The exact cause of the events could not be determined.Sem analysis revealed evidence of elongations and plastic deformations associated with material tensile overload.This was during a procedure to treat two vessels which were stenosed and calcified.Therefore, based on the information available for review, it is assumed the device was induced to a tensile force that exceeded the material yield strength prior to the separation.Vessel characteristics, along with procedural factors and handling of the catheter described contributed to the damages reported.According to the instructions for use, which is not intended as a mitigation of risk, ¿stent placement: if resistance is encountered at any time during the insertion procedure, do not force passage.Resistance may cause damage to the stent or vessel.Carefully withdraw the stent system without deploying the stent.If resistance is felt when beginning deployment, do not force deployment.Carefully withdraw the stent system without deploying the stent.Insertion of guiding sheath or guide catheter and guidewire: a.Access the treatment site utilizing the appropriate accessory equipment compatible with the 6f (2.0 mm) delivery system.B.Place a 6f guiding sheath of an appropriate length (110cm for a 150cm long device and a 135cm for a 190cm long device) with an internal diameter of at least 2.2 mm.C.A brite tip radianztm guiding sheath is highly recommended.D.Place an.018¿ (0.46 mm) guidewire of sufficient length across the lesion to be stented via the guiding sheath or guide catheter.Introduction of stent delivery system: a.Ensure the safety lock is still in place.B.Advance the device over the guidewire through the hemostatic valve and guiding sheath to the lesion site.Note: if resistance is met during delivery system introduction, the system should be withdrawn, and another system should be used.The handle should not be rotated during insertion or removal.Caution: always use a guiding sheath for the implant procedure to protect puncture site.A guiding sheath of a 6f (2.0 mm) or larger size is recommended.Slack removal: a.Advance the stent delivery system past the lesion site.B.Pull back the stent delivery system until the radiopaque stent markers (leading and trailing ends) move in position so that they are proximal and distal to the target lesion site.C.Ensure the device outside the patient remains flat and straight.Caution: slack in the catheter shaft, either outside or inside the patient, may result in deploying the stent beyond the target lesion site.Post-deployment stent dilatation: a.While using fluoroscopy, withdraw the entire delivery system as one unit, over the guidewire, into the guiding sheath and out of the body.Remove the delivery device from the guidewire.Do not rotate the handle during withdrawal.Note: the guide wire lumen will be exposed upon removal of the device from the sheath.Be sure that distal tip is visible outside of the hemostasis valve before attempting to grasp the guide wire.B.Using fluoroscopy, visualize the stent to verify full deployment.If incomplete expansion exists within the stent at any point along the lesion, post deployment balloon dilatation (standard pta technique) can be performed.Note: only areas within the stent length should receive post-deployment balloon dilatation.C.Select an appropriate size pta balloon catheter and dilate the lesion with conventional technique.The inflation diameter of the pta balloon used for post-dilatation should approximate the diameter of the reference vessel.Remove the pta balloon from the patient.¿ the information available does not suggest a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
|