|
Product complaint # (b)(4).Information regarding patient weight, height, race, and ethnicity was not reported.Section d2b ¿ procode: krd/hcg the device was discarded; therefore, no further investigation can be performed.A manufacturing record evaluation was performed for the finished device 31243967 number, and no non-conformances related to the malfunction were identified.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the manufacturing record evaluation, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be determined; however, there are circumstances of the procedure that may have contributed to the reported failure.The instructions for use (ifu) contain the following recommendations: 1.Remove the detacher from the protective packaging and place it within the sterile field.The detacher is packaged separately for single patient use only.2.Confirm again under fluoroscopy that the coil alignment marker of the delivery tube creates a ¿t¿ with the proximal marker of the microcatheter.3.Verify that the rhv is firmly locked around the delivery tube to ensure that the coil does not move during connection process.Ensure the delivery tube is straight between the rhv and the detacher.4.Hold the delivery tube approximately 3 inches from proximal end and bring the detacher over the delivery tube until a firm stop is felt.5.Maintain the delivery tube position within the detacher and use a thumb to pull back the slider on the detacher.6.Continue to pull on the slider until end of travel or a click is heard.Gently release the slider until it returns to starting position or a second click is heard.Remove detacher by slowly withdrawing it from the delivery tube.Note: detacher must remain in sterile field until the procedure is completed.7.Fluoroscopically verify that the coil is detached by gently pulling back the delivery tube approximately 1cm.If the coil has detached, remove the delivery tube from the microcatheter.8.If the coil has not detached, repeat steps 3-7.If 2 unsuccessful detachments occur, discard the detacher and proceed to the manual break section.9.Repeat steps 1-8 until the procedure is completed.Discard the detacher.Caution: always perform fluoroscopic verification of detachment prior to withdrawing the delivery system fully.Failure to do so may lead to an embolic complication.10.After detaching the coil, remove the delivery tube from the microcatheter and discard.11.Repeat the above sequence for all additional coils until the procedure is complete.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.This is one of three products involved with the complaint and the associated manufacturer report numbers are 3008114965-2024-00440 and 3008114965-2024-00441.
|
|
As reported by the field, during a coil embolization to the right carotid terminus, the physician had detachment issues with a uniform xl 10mm x 17.1 cm coil (fcx181017, 31243967).The mechanical detachment handle (mdh1, 102109430) did not break.He was able to get it to detach by pushing the coil pusher rapidly in and out.The next coil was a freeform 9mm x 25 cm (scr120925, 31173116), the handle detachment failed, and he then tried the manual break, this one detached.Then a freeform 6mm x 20 cm coil (scr120620, 31058141) was used.The handle failed and so did the manual break.He tried to make it detach by pushing the coil pusher in and out in hopes it would release.The physician then tried to remove the coil, while pulling it out of the aneurysm the coil detached in the echelon 10 microcatheter (mc).He was able to push the coil back into the aneurysm with a microwire.Additional event information was received on 16-apr-2024 indicating that the uniform xl 10mm x 17.1 cm coil was the first coil to be used.There was no resistance while pushing the coils through the microcatheter; there was no difficulty positioning the microcatheter into the aneurysm however there was difficulty getting up through the arch to get to the aneurysm.The vessel was very tortuous.There was no injury to the patient.The procedure was prolonged for approximately 15 to 20 minutes to the event, the delay was not clinically significant.
|